TC-3 Temperature Controller User’s Manual · Page 15
Menu Description Factory
Default
R-High Lim R-High Lim is the sensor high limit. If
operating in R Mode and the actual
sensor measurement exceeds this
value, the output will be turned off. R
limits are not monitored in Ite Mode.
Limit monitoring can also be disabled in
software (see TEC:ENAB:OUTOFF
command).
45kΩ
Tol Time Tolerance time is the amount of time, in
seconds, that the actual temperature
must be within the set point
temperature +/- the Tol Temp value for
the unit to be considered in tolerance.
5 seconds
Tol Temp Tolerance temperature is a temperature
band (in °C) around the set point
temperature. When the actual
temperature is within this band for
longer than the Tol Time setting, then
the unit is considered to be in tolerance.
0.1°C
H/C Mode This selects the heating and/or cooling
mode of the TC-3. See the section
below titled “Resistive Heaters and
Heat/Cool Only Modes” for more
information.
Heat/Cool
Ext Fan This selects the voltage for the auxiliary
fan power supply. See the External Fan
Control section below for more
information.
Off
Ext Fan Pwr When Ext Fan is set to Custom, this
sets the specific fan voltage from 4.0 to
12.0 volts. See the External Fan Control
section below for more information.
12.0
Ext Fan
Mode
Controls when the fan operates. See the
External Fan Control section below for
more information.
Auto
Ext Fan Off When Ext Fan Mode is set to Delay,
this setting defines the number of
minutes to delay turning off the fan
after the TEC output has been turned
off. See the External Fan Control section
below for more information.
5 minutes