TC-3 Temperature Controller User’s Manual · Page 7
and current begin driving the fixture to the set point you have chosen.
Depending on the thermal size of your fixture, it may take seconds or
several minutes to reach the set point.
If you notice the temperature is oscillating around the set point and not
stabilizing, you may need to adjust the Gain setting in the menu. You
can use the AutoTune feature to automatically calculate the best PID
values, or select from a set of eight factory preset values that typically
cover most applications. To use the factory gains, if the temperature is
quickly jumping up and down, the Gain will typically need to be
reduced. If the temperature is slowly moving up and down, try a higher
Gain. You may need to experiment with several gain settings to find
the ideal value, and for even finer control, you can set the Gain to PID
and directly set the PID control values.
It’s that simple. For more detailed operating and installation
instructions, read on.
Installation
Installation of the TC-3 is very straightforward, as the quick start
section above illustrated. This section will provide additional details
and considerations for installing your TC-3.
After unpacking the unit, make sure all packing materials have been
removed and nothing obscures the ventilation ports on the back and
bottom of the unit.
Changing the Voltage Selection
Before powering on the unit, ensure that the voltage selection on the
IPC is set correctly. Improper voltage selection can easily damage the
unit. Changing the voltage selection requires that you remove the
voltage selection module from the IPC.
Remove the power cord from the unit. Using a small, flat-blade
screwdriver, insert the tip of the screwdriver into the two small
openings above and below the voltage indication to release the voltage
selection module from the IPC, as shown in the picture below.